| Model Number |
87011 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An evaluation of the pelvic screw is not possible.The explanted device has been discarded by the user facility.Neither the identifying part and/or lot number have been provided.The second screwdriver reported is of the same product part number but contains a separate manufacturing lot number : 7405101r1 manufactured 3/2/2015.Evaluation of the returned drivers found that the distal tips of both instruments had fractured and became separated.Inspection under magnification revealed circular fracture lines indicative of a torsional overload.The detached tips were both recovered from the patient and discarded by the user facility.
|
| |
|
Event or Problem Description
|
|
Pelvic screw broke near head about 90mm in.Two screwdrivers were broken during insertion and removal.Ez out eventually removed the broken screw.Event caused an hour delay in the case.
|
| |
|