MAUDE Adverse Event Report: SMITH AND NEPHEW, INC. JOURNEY TIBIA BASE NP RT SZ 6; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH AND NEPHEW, INC. JOURNEY TIBIA BASE NP RT SZ 6; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 74022216
Health Effect - Clinical Codes Rupture (2208); No Code Available (3191)
Date of Event 06/29/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a quadricep tendon rupture.Action taken: surgery/procedure - revision with poly change, debridement and quad repair.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A clinical evaluation was conducted and based on the available information, the root cause of this revision was as stated¿¿the fall at home¿.This was not a device failure.Follow-up information has been provided; the patient has a re-tear of the quadriceps (in a radiology report).A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY TIBIA BASE NP RT SZ 6
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH AND NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7845916
Report Number1020279-2018-01703
Device Sequence Number19118567
Product Code JWH
UDI-Device Identifier03596010554307
UDI-Public03596010554307
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K042515
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number74022216
Device Lot Number17MJM14820
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/16/2018
Supplement Date Received by Manufacturer08/16/2018
Initial Report FDA Received Date09/05/2018
Supplement Report FDA Received Date01/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-