MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR LINERS Back to Search Results
Catalog Number UNK HIP ACETABULAR LINER
Device Problems Material Separation; Device Dislodged or Dislocated; Naturally Worn; Separation Problem
Event Date 08/13/2018
Event Type  Injury  
Manufacturer Narrative

(b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. Concomitant medical products: unknown femoral head.

 
Event Description

The patient was revised to address metallosis. The acetabular cup is worn and has a hole in it. The liner is intact but is worn out. The acetabular screw broke off inside the synthes trephine. One trephine broke. The screw was removed after 45 minutes. Osteolysis was present as well as loosening of the cup at the bone to implant interface. The summit stem was intact so it was not removed. Doi: 2009, dor: (b)(6) 2018, left hip.

 
Manufacturer Narrative

Product complaint #: (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN HIP ACETABULAR LINERS
Type of DeviceHIP ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester , IN 19380-0988
6103142063
MDR Report Key7845944
Report Number1818910-2018-68447
Device Sequence Number1
Product CodeLZO
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 08/09/2018
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received09/05/2018
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue NumberUNK HIP ACETABULAR LINER
Device LOT NumberUNKNOWN
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/09/2019
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 09/05/2018 Patient Sequence Number: 1
-
-