The patient was revised to address metallosis.The acetabular cup is worn and has a hole in it.The liner is intact but is worn out.The acetabular screw broke off inside the synthes trephine.One trephine broke.The screw was removed after 45 minutes.Osteolysis was present as well as loosening of the cup at the bone to implant interface.The summit stem was intact so it was not removed.Doi: 2009, dor: (b)(6) 2018, left hip.
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # = > (b)(4).Investigation summary = > no device associated with this report was received for examination.Examination of the provided x-rays and photographs confirmed the reported event.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot code(s) was not provided.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, will be filed as appropriate.
|