| Catalog Number |
136536000 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Host-Tissue Reaction (1297); Pain (1994); Loss of Range of Motion (2032); Tissue Damage (2104); Distress (2329); Discomfort (2330); Injury (2348); Limited Mobility Of The Implanted Joint (2671); Test Result (2695); Not Applicable (3189)
|
| Date of Event |
01/11/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Reporter is an attorney.
|
| |
|
Event or Problem Description
|
|
Litigation alleges friction and wear between the cobalt-chromium metal head and cobalt-chromium metal liner caused large amounts of toxic metal ions and particles released in the blood, tissue and bone surrounding the implant.It also alleges that the plaintiff experienced injury, pain, discomfort, loss of mobility, loss of range of motion, and emotional distress.The surgeon noted presence of pseudotumor, and blood work showed elevated levels of metal ions prior to the removal of the prosthesis.Doi: (b)(6) 2011 - dor: (b)(6) 2017; left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
Ppf alleges pseudotumor.Pfs alleges metal allergy.After review of medical records, patient was revised to address failed metal on metal total hip arthroplasty with pseudotumor.Revision notes reported a significant scar tissue was encountered posteriorly consistent with a history of posterior approach.There was also a mild corrosion noted of the neck upon disimpacted from the head.
|
| |
|