• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC TRILOGY 02; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC TRILOGY 02; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1040004
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/13/2018
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to the manufacturer for routine preventative maintenance.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, a "service required" code was found in the ventilator's downloaded error log.The device's oxygen blending module board and interface board was replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 02
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC
1001 murry ridge lane
murrysville PA 15668
MDR Report Key7846242
Report Number2518422-2018-02061
Device Sequence Number1573579
Product Code CBK
UDI-Device Identifier00606959033887
UDI-Public00606959033887
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K093905
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1040004
Device Catalogue Number1040004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/13/2018
Initial Report FDA Received Date09/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-