Model Number LF1520
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the device had no sealing during surgery.
 
Additional Manufacturer Narrative
One device was received for evaluation.The returned product met specification as received.Visual inspection found no defects.Functional testing was performed with the returned device on porcine kidney tissue.Multiple seals on various size vessels were made and a full thermal effect was visually verified.The device was activated while pressing the button and with the rotation knob in vario us positions to detect any problematic activation issues.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The investigation found the device to function normally and within specifications.The instructions for use (ifu) states, keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Correction: event was notified to the manufacturer on (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key7846502
Report Number1717344-2018-01169
Device Sequence Number1313691
Product Code GEI
UDI-Device Identifier20884521125527
UDI-Public20884521125527
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K092879
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2021
Device Model NumberLF1520
Device Catalogue NumberLF1520
Device Lot Number60530129X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2018
Initial Date Received by Manufacturer 06/18/2018
Supplement Date Received by Manufacturer08/29/2018
10/12/2018
Initial Report FDA Received Date09/05/2018
Supplement Report FDA Received Date09/14/2018
11/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-