| Catalog Number |
3341459 |
| Medical Device Problem Code |
Fungus in Device Environment (2316)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A device history record review is in progress.Once completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient was going to undergo breast implantation procedure of unknown type with a gel mentor memoryshape breast implant 615cc and it was discovered that the implant was received with mold inside the box.There was no patient contact with implant.
|
| |
|
Additional Manufacturer Narrative
|
|
On 11/26/18, it was reported to mentor that the device history record (dhr) for the lot number 7308224 has been reviewed and it was verified that the device was manufactured in accordance with documented specifications and procedures.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 3/25/2019, it was reported to mentor that the device analysis has been completed.Multiple attempts were performed to request the device, however no response was received.Since the product was not returned for evaluation no product failure analysis can be conducted, and no determination of possible contributing factors could be made.No related complaints were found for this lot number.If the product is received in the future the investigation will continue.Manufacturing controls: there are multiple controls within the manufacturing process.At multiple points the devices are processed to ensure any foreign matter, particulates or contaminants are removed or 100% visually inspected for any defects including foreign matter or particulates and subsequently rejected when identified.Review applicable fmeas: the application failure modes and effects analysis (afmea) was reviewed for the defect in question (foreign matter/mold) with the severity being ¿s3¿ and the hazard risk was determined to be as low as possible (alap), with no further action required.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2019, it was reported to mentor that the manufacturing record evaluation (mre) was performed for the finished device 7308224 number, and no non-conformance's related to the reported complaint condition were identified.Manufacturer¿s reference number: (b)(4).
|
| |
|