| Model Number |
8886614921 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Tissue Damage (2104); No Code Available (3191)
|
| Date of Event |
08/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a proximal pancreaticoduodenectomy, while the device was used for anastomosis of biliary duct and anastomosis of jejunum, the device needle was detached from the thread, to resolve the issue the surgeon repeated the anastomoses and used another device to complete the case.The surgical time was extended more than 30 minutes and there was a tissue damage due to the device problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection and functional testing confirmed there were no abnormalities that would have caused or contributed to the reported condition.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a proximal pancreaticoduodenectomy, while the device was used for anastomosis of biliary duct and anastomosis of jejunum, the device needle was detached from the thread, to resolve the issue the surgeon repeated the anastomoses and used another device to complete the case.The surgical time was extended more than 30 minutes and there was a tissue damage due to the device problem.
|
| |
|
Search Alerts/Recalls
|