|
(b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual and functional inspection was performed on the returned device.The reported balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other incidents.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.It is likely that the reported balloon rupture occurred due to interaction with anatomy or associated devices.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|