Catalog Number CAT3
Medical Device Problem Codes Unraveled Material (1664); Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the cat3 was fractured approximately 101.7 cm from the hub.The cat3 was damaged and, therefore, could not be functionally tested.Conclusions: evaluation of the returned cat3 revealed that the catheter was fractured.If the device is forcefully retracted out of its packaging hoop at an extreme angle, damage such as a fracture may occur.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a thrombectomy procedure, the hospital staff found that the indigo system cat3 aspiration catheter (cat3) was unraveling upon removal from the packaging.The damage to the cat3 was found prior to use.Therefore, the cat3 was not used in the procedure.The procedure was completed using a new cat3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM CAT3 ASPIRATION CATHETER
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7847355
Report Number3005168196-2018-01719
Device Sequence Number1338152
Product Code DXE
UDI-Device Identifier00814548016085
UDI-Public00814548016085
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/06/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/14/2020
Device Catalogue NumberCAT3
Device Lot NumberF79544
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2018
Initial Report FDA Received Date09/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-