Catalog Number 4002C1450
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The embolization coil was intact with its pusher assembly and had offset coil winds.Conclusions: evaluation of the returned pc400 revealed that the embolization coil winds were offset.If the embolization coil is forcefully advanced against resistance, damage such as offset coil winds may occur.If the device is subsequently attempted to be advanced after the coil winds are offset, resistance may be experienced.During the functional test, resistance was encountered while advancing the pc400 out of its introducer sheath and the device could not be advanced any further.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a ruptured anterior communicating artery aneurysm (acom) using penumbra coils 400 (pc400s).During the procedure, the physician had difficulty advancing a pc400 through a px slim delivery microcatheter (px slim); therefore, it was removed.It was reported that the pc400 would not advance more than half way in the px slim.The procedure was completed using additional pc400s and the same px slim.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7847671
Report Number3005168196-2018-01720
Device Sequence Number1313626
Product Code HCG
UDI-Device Identifier00814548010687
UDI-Public00814548010687
Combination Product (Y/N)Y
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/06/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/15/2023
Device Catalogue Number4002C1450
Device Lot NumberF65335
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2018
Initial Report FDA Received Date09/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR