| Catalog Number |
4002C0610 |
| Medical Device Problem Codes |
Difficult or Delayed Activation (2577); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2018-01725.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the internal iliac artery (iia) using penumbra coils 400 (pc400s).During the procedure, the physician advanced a pc400 in the target vessel using a px slim delivery microcatheter (px slim) and used a penumbra coil 400 detachment handle (handle) to detach the pc400.Although the detachment zone was cut off, the physician pulled the pc400 pusher assembly and observed that the pc400 was still attached to its pusher assembly; therefore, the physician decided to remove the pc400.However, while withdrawing the pc400 and the px slim together, the pc400 unintentionally detached inside the non-penumbra guiding catheter.The physician then withdrew the pc400 and the guiding catheter together.The procedure was completed using a new pc400, the same handle, px slim and guiding catheter.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Results: the pet lock was separated on the proximal end of the pusher assembly.The pull wire was retracted out of the distal detachment tip (ddt) and the ddt was undamaged.The embolization coil was undamaged and detached from the pusher assembly.Conclusion: evaluation of the returned pc400 confirmed a device with a separated pet lock and a detached embolization coil.The reported complaint mentioned that the handle was used to detach the pc400.A separated pet lock indicates a pull force was applied to the proximal end of the pusher assembly, typically in an attempt to detach the coil with a handle.A separated pet lock indicates the pull wire is retracted from the ddt.The root cause of the difficulty detaching the embolization coil could not be determined.If the pull wire is retracted from the ddt and the pusher assembly is retracted, the embolization coil will likely detach.The non-penumbra catheter was not returned for evaluation.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Search Alerts/Recalls
|