| Catalog Number |
48750110 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during lateral cage insertion, the cage broke.Another cage of the same size was used to successfully complete the procedure.The procedure was at the l2-l3 level ans insertion angle was noted difficult.All broken fragments were retrieved and no reports of adverse consequences to the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that during lateral cage insertion, the cage broke.Another cage of the same size was used to successfully complete the procedure.The procedure was at the l2-l3 level ans insertion angle was noted difficult.All broken fragments were retrieved and no reports of adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection, product history review, complaint history review, nc/capa history review, labelling review.Result: the reported event was confirmed via email correspondence with the sales rep.Upon device inspection, the device found to have its portion where the inserter is inserted fractured off.No manufacturing related issues were found upon device history review.Conclusion: based on the information provided, the probable root cause of the reported event is difficult angle for insertion of the cage and excessive cantilever force applied during insertion.
|
| |
|