• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESHAPE LIFE SCIENCES MAESTRO RECHARGEABLE SYSTEM; RECHARGEABLE NEUROREGULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESHAPE LIFE SCIENCES MAESTRO RECHARGEABLE SYSTEM; RECHARGEABLE NEUROREGULATOR Back to Search Results
Model Number 2002
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code Therapeutic Response, Decreased (2271)
Date of Event 08/01/2018
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2018, patient (b)(6) at the university (b)(6) was unable to charge the device and had a flashing red light on the mc.The reshape lifesciences field specialist connected her to the cp and alarm codes #4 and #25 were observed.These alarms were cleared and the flashing red light disappeared.The cp continued to ask to start a charge as the rnr battery was extremely low.After about 15 minutes of charging using the patients mc, therapy was able to be adjusted and a live test performed.On-demand impedance measurement was performed and the results were as follows: 934, 1033, 677.Before the patient left, another on-demand impedance test was performed, the readings were as follows: 821, 1062, 0.A red flashing light was observed on the mc.The patient was hooked up to the cp to clear the red light and was sent home.On (b)(6) 2018, the patient indicated they received another flashing red light.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAESTRO RECHARGEABLE SYSTEM
Common Device Name
RECHARGEABLE NEUROREGULATOR
Manufacturer (Section D)
RESHAPE LIFE SCIENCES
2800 patton road
saint paul MN 55113
MDR Report Key7848028
Report Number3005025697-2018-00009
Device Sequence Number1293764
Product Code PIM
UDI-Device Identifier00857334004262
UDI-Public00857334004262
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P130019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number2002
Device Catalogue Number2002
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2018
Initial Report FDA Received Date09/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-