• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Codes Failure to Deliver Shock/Stimulation (1133); Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio-control then replaced the main keypad assembly and after observing proper device operation through functional and performance testing the unit was returned to the customer for use.Based on the information available, physio-control has determined that it¿s reasonable to conclude that the likely cause of the reported issue was that the shock switch located on the main keypad assembly was out of tolerance due to excessive resistance.When this issue occurs, an event code will be logged in the device's memory following a failure of the device to deliver defibrillation energy.
 
Event or Problem Description
The customer contacted physio-control to report that their device had failed a user test.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that it had logged an event code in the memory which is related to a potential failure of the device to deliver defibrillation energy.
 
Additional Manufacturer Narrative
Initial report sent to fda of the initial mdr should have indicated ¿ no.
 
Event or Problem Description
The customer contacted physio-control to report that their device had failed a user test.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that it had logged an event code in the memory which is related to a potential failure of the device to deliver defibrillation energy.
 
Additional Manufacturer Narrative
This product complaint requires a supplemental mdr to document that field action number 301587611/18/2019-001c is relevant to the reported issue.
 
Event or Problem Description
The customer contacted physio-control to report that their device had failed a user test.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that it had logged an event code in the memory which is related to a potential failure of the device to deliver defibrillation energy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7848483
Report Number0003015876-2018-01461
Device Sequence Number1574224
Product Code MKJ
UDI-Device Identifier00883873979419
UDI-Public00883873979419
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-000056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2018
Initial Date Received by Manufacturer 08/13/2018
Supplement Date Received by Manufacturer11/18/2019
04/02/2020
Initial Report FDA Received Date09/05/2018
Supplement Report FDA Received Date11/21/2019
04/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3015876-11/18/2019-001C
Patient Sequence Number1
-
-