Catalog Number 5573-2E-3602
Health Effect - Clinical Codes Fall (1848); Pain (1994)
Date of Event 08/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed, any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that patient's right shoulder was revised after patient suffered several falls and complained of pain.The glenoid baseplate was found to have loosened.Surgeon decided to use bone allograft in the glenoid and a humeral head on the existing stem.
 
Additional Manufacturer Narrative
The reported event that glenosphere - 36mm dia x 2mm thk 2mm eccentric was alleged of 'implant - dislocated' could be confirmed.Based on investigation, the root cause was attributed to be patient related.The failure was caused by as it was reported that the patient suffered several falls.The device inspection revealed the following: some damages and deformations are clearly visible on the returned implants.The close up views (done by the microscope) showing the deformations of the screw holes in the baseplate, which clearly indicate that the force during the patient falls must have been immense.Which resulted in the dislocation of the 3 returned screws.It is also quit evident that the glenosphere is badly worn, which indicates, as the patient had several falls, that the implant positioning afterwards was not correct anymore.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Note that we are still waiting for a statement of our medical expert.If any further information is provided which differ from the current determination, the investigation report will be updated.
 
Event or Problem Description
It was reported that patient's right shoulder was revised after patient suffered several falls and complained of pain.The glenoid baseplate was found to have loosened.Surgeon decided to use bone allograft in the glenoid and a humeral head on the existing stem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLENOSPHERE - 36MM DIA X 2MM THK 2MM ECCENTRIC
Common Device Name
SHOULDER JOINT METAL/POLYMER SEMI-CONSTRAINED CEMENTED PROSTHESIS.
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key7848582
Report Number0008031020-2018-00580
Device Sequence Number16478188
Product Code KWS
UDI-Device Identifier07613327098914
UDI-Public07613327098914
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K130895
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/24/2020
Device Catalogue Number5573-2E-3602
Device Lot NumberNA5YTY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2018
Initial Date Received by Manufacturer 08/08/2018
Supplement Date Received by Manufacturer11/22/2018
Initial Report FDA Received Date09/05/2018
Supplement Report FDA Received Date12/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight118
-
-