MAUDE Adverse Event Report: NEODENT - JJGC S.A. CM DRIVE IMPLANT 5.0X16; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. CM DRIVE IMPLANT 5.0X16; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 109.687
Medical Device Problem Codes Failure to Osseointegrate (1863); Loss of Osseointegration (2408)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994)
Date of Event 06/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
 
Event or Problem Description
The clinician reports that more than six months after the dental implant and bridge were placed in ada site 4 in the mouth, the implant lost ntegration in type ii bone.Clinician reports trauma, pain and mobility.The site was grafted.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.
 
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.After the evaluation, the event was considered as osseointegration failure.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.
 
Event or Problem Description
The clinician reports that more than six months after the dental implant and bridge were placed in ada site 4 in the mouth, the implant lost integration in type ii bone.Clinician reports trauma, pain and mobility.The site was grafted.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM DRIVE IMPLANT 5.0X16
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key7848592
Report Number3008261720-2018-04464
Device Sequence Number16467194
Product Code DZE
UDI-Device Identifier07899878023487
UDI-Public07899878023487
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K123022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 09/05/2018
Report Date (Section F) 08/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/17/2021
Device Catalogue Number109.687
Device Lot Number86893I
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date11/16/2018
Device Age1 MO
Event Location Other
Date Report to Manufacturer08/09/2018
Initial Date Received by Manufacturer 08/09/2018
Supplement Date Received by Manufacturer10/24/2018
Initial Report FDA Received Date09/05/2018
Supplement Report FDA Received Date11/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BRIDGE PLACEMENT; BRIDGE PLACEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
-
-