| Catalog Number |
109.687 |
| Medical Device Problem Codes |
Failure to Osseointegrate (1863); Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994)
|
| Date of Event |
06/06/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report
on behalf of neodent - jjgc.
|
| |
|
Event or Problem Description
|
|
The clinician reports that more than six months after the dental implant and bridge were placed in ada site 4 in the mouth, the implant lost ntegration in type ii bone.Clinician reports trauma, pain and mobility.The site was grafted.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.After the evaluation, the event was considered as osseointegration failure.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.
|
| |
|
Event or Problem Description
|
|
The clinician reports that more than six months after the dental implant and bridge were placed in ada site 4 in the mouth, the implant lost integration in type ii bone.Clinician reports trauma, pain and mobility.The site was grafted.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.(b)(4).
|
| |
|