• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Catalog Number 8590855
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: microscopic and optical examination revealed thread crest/flank damage; this damage appears to have initiated near the start of the thread, and is evident around the damaged portion of the thread.Functional evaluation with a sample mas head found the set screw was still able to be fully engaged.The above observations are consistent with misalignment of the mas and set screw threads during construct assembly.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported intra-operatively the screw was slipped during surgery.The device come in contact with the patient.No patient complications were reported due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key7849825
Report Number1030489-2018-01211
Device Sequence Number1263692
Product Code KWP
UDI-Device Identifier00721902365263
UDI-Public00721902365263
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K981676
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8590855
Device Lot NumberH5339778
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2018
Initial Report FDA Received Date09/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
-
-