The complaint device was received and evaluated.Visual observation confirms the weld breakage reported as the sheath end is completely separated from the t-handle on the device.The device also exhibits wear as would be typical with excessive abuse of the device over repeated uses.This excessive abuse would be typical of the separation of the two ends from each other as is noted in this complaint.It was noted that the design of this device was changed making it a single piece design, thus eliminating the weld.Therefore, it can be determined that this design change has corrected the separation failure mode.Furthermore, no lot numbers were provided which precludes conducting a batch history review or a lot specific search in the complaints handling system.Based on the overall complaint rate and implemented design changes, at this point in time, no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4)-incomplete.The lot number is currently unavailable.This report is being filed from the etq complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
|
It was reported by the sales rep that during an anterior cruciate ligament (acl) surgical procedure, it was observed that the top part came on the customer's biointrafix sheath trial 7-8mm off when removing but the bottom part stayed in.The sales rep stated that the surgeon had to mallet the bottom part out.The sales rep reported that the surgeon had completed the procedure with the device before the failure occurred.The sales rep reported that there were no patient consequences or delays in the procedure.The sales rep stated that the device broke outside the patient and there was no debris in the patient.The sales rep could not provide a lot number for the device but stated that the device has seen heavy use in the field.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|