ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Structural Problem (2506)
Health Effect - Clinical Codes Hyperglycemia (1905); Urinary Frequency / Polyuria (2275); Lethargy (2560)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.
 
Event or Problem Description
On (b)(6) 2018, a reporter contacted animas alleging that a patient had experienced a hyperglycemic event while on the pump.The reporter stated that the patient had a blood glucose of over 500 mg/dl with symptoms of drowsiness and excess urination.The patient did not receive any treatment above and beyond the normal course of diabetes management.The reporter stated that the patient experienced the blood glucose excursion after the pump stopped working.The reporter alleged that this was related to a power issue with the pump.The battery cap was found to be damaged.This complaint is being reported based on the allegation that the patient experienced a hyperglycemic event while on the pump due to a power issue.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 18-oct-2018 with the following findings: the battery compartment was found to be cracked.A test cap was unable to fully tighten to the pump, but was able to power it on.There was evidence of moisture corrosion visible in the battery compartment.The pump's current draws were within specifications.The pump passed a delivery accuracy test with no issues.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key7850149
Report Number2531779-2018-16120
Device Sequence Number19148292
Product Code OYC
UDI-Device Identifier10840406100198
UDI-Public10840406100198
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K080639
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2018
Was the Report Sent to FDA? Yes
Device Age26 MO
Initial Date Received by Manufacturer 08/10/2018
Supplement Date Received by Manufacturer08/10/2018
Initial Report FDA Received Date09/06/2018
Supplement Report FDA Received Date10/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient Age47 YR
-
-