Model Number N/A
Device Problem Gas/Air Leak (2946)
Patient Problems Hematoma (1884); No Code Available (3191)
Event Date 08/08/2018
Event Type  Injury  
Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.(b)(4).The actual device was not returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.The production lot number was not provided by the user facility, which prevented a meaningful review of the device history record.
 
Event Description
The user facility reported that the tr band device leaked air.The ccl had placed a tr band post radial procedure and the balloon on the band leaked out air overtime and created a hematoma on the patient.It was reported that this was the first tr band this happened with and the physician placed the band on the patient.The procedure was successful and there was no blood loss from the patient due to the tr band issue.Additional information was received on september 5, 2018.Once the nurse saw the patient had a hematoma, they removed the band and held pressure to evacuate the hematoma.
 
Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.With no device return the exact cause of the reported event cannot be definitively determined based on the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TR BAND
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer (Section G)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer Contact
mark vornheder
950 elkton blvd.
elkton, MD 21921
8002837866
MDR Report Key7850192
MDR Text Key119337022
Report Number1118880-2018-00119
Device Sequence Number1
Product Code DXC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberTRB24-REG
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2018
Initial Date FDA Received09/06/2018
Supplement Dates Manufacturer Received09/11/2018
Supplement Dates FDA Received09/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SLENDER SHEATH
Patient Outcome(s) Required Intervention;
-
-