• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Moisture Damage (1405); Structural Problem (2506)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to the manufacturer and investigation completed 27-aug-2018.On examination, the pump was confirmed to have a crack in the battery compartment.Moisture corrosion was also noted inside the battery compartment.A battery cap was not returned with the pump for investigation; a test cap was used to complete the investigation.A test cap fit properly to the pump; however, the pump failed to power on for investigation.The pump was unresponsive; there was no vibration or audio tone and the display screen remained blank.Leak testing confirmed moisture ingress into the battery compartment at the site of the battery compartment crack.There was no damage, defect or contamination of the pump's interior components.Investigation duplicated the complaint.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging a power (damage) issue.The reporter alleged the battery compartment was cracked and the pump experienced an intermittent power issue.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key7850224
Report Number2531779-2018-16124
Device Sequence Number1939740
Product Code OYC
UDI-Device Identifier10840406100341
UDI-Public10840406100341
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2018
Was the Report Sent to FDA? Yes
Device Age25 MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2018
Initial Report FDA Received Date09/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age20 YR
Patient Weight80
-
-