| Catalog Number |
01.32.6540 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 04 september 2018.Lot 180587: (b)(4) items manufactured and released on 03 april 2018.Expiration date: -03-20.2023.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
During final tightening a screw head broke off.The surgeon removed the screw head, but the threads remained in the patient.The surgeon used a secondary screw completing successfully the surgery.
|
| |
|
Search Alerts/Recalls
|