| Catalog Number |
9733498 |
| Medical Device Problem Codes |
Imprecision (1307); Material Deformation (2976)
|
| Health Effect - Clinical Codes |
Spinal Column Injury (2081); Tissue Damage (2104)
|
| Date of Event |
08/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturing date was unavailable at the time of filing.No parts have been returned to the manufacturer for analysis.Part not returned.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system.It was reported that during the navigate task of a sacroiliac and thoracolumbar procedure, there was an alleged inaccuracy due to a damaged navigated pedicle access needle (pac).The navigated pac needle was used for a l2-l4 posterior fixation "mis voyager procedure".The pac needle registered using a passive spine frame.The pac needle was used to find the appropriate guidewire position.After insertion of all 6 guidewires the navigated cannulated tab driver was used.Having the guidewires in place and following the acquired path with the navigated tab driver it was alleged that the path was 2mm off the actual target path.After eliminating possible errors - the nav tab and nav driver showed the same path - a new pac needle was used.That showed that the first pac needle was faulty.The site examined the faulty pac needle and it could be seen that the cannula was siting loosely in the handle - which explained why the surgeon was able to register the pac needle - not putting pressure on it.However as soon as the pac needle was within tissue the surgeon was putting pressure on it and therefore showed a different path on the navigation system than its actual path.This incident added approx 2 hours to hour case, which normally would have only taken 1.5 hours due to the fact that the surgeon had to re-position its guidewires and two already inserted screws.The final result was good, there was no reported impact on patient outcome.
|
| |
|
Search Alerts/Recalls
|