| Model Number |
E SERIES |
| Medical Device Problem Code |
Failure to Charge (1085)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed to charge.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device.The reported malfunction was duplicated and attributed to the front panel membrane.The front panel membrane was replaced to resolve the malfunction.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|