| Model Number |
MMT-722NAS |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed motor error and motor position encoder error.Customer's blood glucose level was 131 mg/dl.The customer was able to rewind the insulin pump due to alarm.Customer was advised to discontinue use of the device and revert to a backup plan.The customer was advised that the device would be replaced and agreed to return the product for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement test, rewind, basic occlusion, occlusion, prime and excessive no delivery test.No motor error alarm noted.Motor passed motor test.Unable to verify motor position encoder error alarm in the alarm history due to history file overwritten with displayable history crc check failed on startup alarms.Displayable history crc check failed on startup alarms due to a corrupted history file.
|
| |
|