Catalog Number TCT55
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Is the current patient status known? was there any patient consequence or change in the post-operative care of the patient as a result of the event?.
 
Event or Problem Description
It was reported that during an ileostomy reversal the device was cutting but not stapling.They had to cut more of the anastomis.Case completed with another device of the same product code.
 
Additional Manufacturer Narrative
(b)(4).Batch # r5745m.Device evaluation: the analysis results found that the tct55 device was received with no apparent damage, with no reload present.The device was tested for functionality with a test reload and it fired, cut and formed all the staples as intended.The device fired and cut without any difficulties, the staple line was complete, the cut line was complete and the staples were noted to have the proper b-formed shape.The event described could not be confirmed as the device performed without any difficulties noted.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7851927
Report Number3005075853-2018-12522
Device Sequence Number16635852
Product Code GDW
UDI-Device Identifier20705036013140
UDI-Public20705036013140
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Catalogue NumberTCT55
Device Lot NumberR4058P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2018
Initial Date Received by Manufacturer 08/09/2018
Supplement Date Received by Manufacturer09/27/2018
Initial Report FDA Received Date09/06/2018
Supplement Report FDA Received Date10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-