| Model Number |
5610 |
| Medical Device Problem Code |
Pacemaker Found in Back-Up Mode (1440)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/23/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During a clinic follow up, it was noted that the device went into back up mode.The device was explanted and exchanged to resolve the event and the patient was stable following the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the device image obtained from the pacemaker noted the device was in bvvi mode.The battery voltage and current trends were normal.Functional testing after reloading the product code resulted in normal device characteristics.The reported event of backup vvi mode was confirmed; however, the cause of the reset could not be determined from the device analysis in the laboratory setting.
|
| |
|