• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. PEN NEEDLE; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. PEN NEEDLE; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number PEN NEEDLE
Medical Device Problem Code Failure to Eject (4010)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned product evaluated with no defects found.Most likely underlying root cause: mlc-8: failure to follow instructions.Manufacturer contacted customer on (b)(6) 2018 in a follow-up: customer states that send the product for investigation.No symptoms or medical attention reported.
 
Event or Problem Description
Consumer reported complaint for needle stick due to the pen needle being stuck on the pen.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product storage location is undisclosed.Customer called in states he is unable to take the pen needle of the injector.Customer states he put the clear protective cap and it does not come off when he pulls it off.Customer states he had a needle stick due to the pen needle being stuck on the pen.Customer has a compatible quick pen which is on our approved list.Customer denies the need for medical attention at the time of call.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEN NEEDLE
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7852290
Report Number1000113657-2018-00927
Device Sequence Number13878884
Product Code FMI
UDI-Device Identifier00021292009069
UDI-Public(01)00021292009069(17)220701(10)X56S1
Combination Product (Y/N)N
Initial Reporter StateWI
PMA/510(K) Number
K143437
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberPEN NEEDLE
Device Catalogue NumberN/A
Device Lot NumberX56S1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2018
Initial Report FDA Received Date09/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-