| Model Number |
M0068404020 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No event date was given, therefore an approximate date of (b)(6) 2018 was used as the event date.(b)(4).The device has been received for analysis; however the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that an accumax 365 laser fiber was used in procedure performed on an unknown date.According to the complainant, during the procedure, it was noted that the laser fiber was broken.There were no serious injury nor adverse patient effects reported as a result of this event.Attempts to obtain additional information regarding this event have been unsuccessful to date.Should additional relevant details become available; a supplement report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
No event date was given, therefore an approximate date of (b)(6) 2018 was used as the event date.(b)(4).The fiber was returned kinked approximately 2 cm from the distal tip.The condition of the returned device confirmed that the fiber was broken.Review and analysis of all available information indicated that the root cause is manufacturing deficiency.A complaint with a cause of manufacturing deficiency indicates that the problems were traced to the manufacturing process.An investigation was completed to address this issue which identified the most probable root cause of the fiber break as handling during the manufacturing process caused by the supplier manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that an accumax 365 laser fiber was used in procedure performed on an unknown date.According to the complainant, during the procedure, it was noted that the laser fiber was broken.There were no serious injury nor adverse patient effects reported as a result of this event.Attempts to obtain additional information regarding this event have been unsuccessful to date.Should additional relevant details become available; a supplement report will be submitted.
|
| |
|