• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 014 PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 014 PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U4150230
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the tibial artery, the pta balloon had incompatibility issues with the guide wire.No further treatment was needed.There was no reported patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the tibial artery, the pta balloon had incompatibility issues with the guide wire.No further treatment was needed.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Three segments of the device and one electronic video were returned and evaluated for evaluation.A visual inspection found a completed detachment in the outer catheter shaft.The detached distal portion of the catheter shaft and balloon was not returned.The inner guidewire lumen was found to be cut from the wire and was returned in two segments.Based on the video review, the device stuck on the guidewire could be confirmed.Therefore, the investigation is confirmed for the reported guidewire incompatibility.The investigation is also confirmed for catheter detachment.It is likely that the issues with the guidewire contributed to the outer catheter detachment.However, the definitive root cause for the identified guidewire issue could not be determined based on the available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 014 PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7852763
Report Number2020394-2018-01662
Device Sequence Number16569826
Product Code LIT
UDI-Device Identifier00801741064425
UDI-Public(01)00801741064425
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K121856
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model NumberU4150230
Device Catalogue NumberU4150230
Device Lot NumberCMCP0178
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2018
Initial Date Received by Manufacturer 08/10/2018
Supplement Date Received by Manufacturer11/28/2018
Initial Report FDA Received Date09/06/2018
Supplement Report FDA Received Date12/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Patient Weight108
-
-