• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION MICROBORE EXTENSION SET, IV CONNECTOR, T; SET, EXTENSION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION MICROBORE EXTENSION SET, IV CONNECTOR, T; SET, EXTENSION, INTRAVASCULAR Back to Search Results
Model Number MZXT9001
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Although requested, the affected product has not been received.A follow up report will be submitted with investigation results should the devices be received for evaluation.
 
Event or Problem Description
The customer reported fluid leaked from the t-connector extension set when attached to the patient's picc line and flushed.There was no report of patient harm.
 
Additional Manufacturer Narrative
The customer¿s report of a leaking t-connector was confirmed.Visual inspection of the set showed no damage or any anomalies.Examination under magnification revealed a thin crack spanning the length of the female luer body.Functional testing confirmed leaking at the attachment of the valve to the female luer.The cause of the leak was determined to be a cracked female luer body.The root cause of the crack was not determined.
 
Event or Problem Description
The customer reported fluid leaked from the t-connector extension set when attached to the patient's picc line and flushed.There was no report of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROBORE EXTENSION SET, IV CONNECTOR, T
Common Device Name
SET, EXTENSION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key7852896
Report Number9616066-2018-01562
Device Sequence Number11259303
Product Code FPA
UDI-Device Identifier10885403240805
UDI-Public10885403240805
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K140831
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMZXT9001
Device Catalogue NumberMZXT9001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/06/2018
Supplement Report FDA Received Date11/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PICC LINE, FLUSH SYRINGE, THERAPY DATE UNK; PICC LINE, FLUSH SYRINGE, THERAPY DATE UNK
-
-