Catalog Number S02020-10
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned, at this time, to the manufacturer for evaluation.A lot history review (lhr) of asbts0143 showed one other similar product complaint(s) from this lot number.Device has not yet been returned for evaluation.
 
Event or Problem Description
It was reported that during the internal inspection, one device unit was with an unidentified colorless material blocking the infusion tube.The device was not used on patient.
 
Event or Problem Description
It was reported that during the internal inspection, one device unit was with an unidentified colorless material blocking the infusion tube.The device was not used on patient.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of material within the extension set tube is confirmed as supplier related.One 20 ga x 1 in miniloc safety infusion set was returned for investigation.The protective sheath was present over the needle.The safety mechanism was not activated.Fluid was observed within the tubing.The fluid observed within the tubing is likely to be silicone oil used in the manufacturing process.The sample was flushed with water using a 12 ml syringe and was found to be patent to infusion.A non-complainant wire was passed through the extension tubing and no obstructions were observed.The fluid observed within the extension tubing appeared to correspond to the material described in the event description and is likely silicone oil; therefore, the complaint is confirmed.The supplier has been notified of this event.A lot history review (lhr) of asbts0143 showed one other similar product complaint(s) from this lot number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINILOC SAFETY INFUSION SET 20G X 1 IN
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7852958
Report Number3006260740-2018-02367
Device Sequence Number1633287
Product Code FPA
UDI-Device Identifier00801741048722
UDI-Public(01)00801741048722
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K050600
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberS02020-10
Device Lot NumberASBTS0143
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer11/13/2018
Initial Report FDA Received Date09/06/2018
Supplement Report FDA Received Date11/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-