• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH HEXAPOD STRUT, LONG HOFFMANN LRF 183MM-295MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH HEXAPOD STRUT, LONG HOFFMANN LRF 183MM-295MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Catalog Number 49350040
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that patient presented to surgeon's office with a broken long strut.This was not a routine followup visit.Causes or contributors to strut breakage currently unknown.A wire was utilized to hold the strut in place until it can be replaced (b)(6) 2018 as the surgeon is away until that date.Rep won't be able to get any additional information until on or after (b)(6) 2018 when he can confer with the surgeon, but expects the device to be available for return after that date.
 
Additional Manufacturer Narrative
The reported event that hexapod strut, long hoffmann lrf 183mm-295mm was alleged of 'implant breakage after surgery' could be confirmed.Based on investigation, the root cause was attributed to be manufacturing and supplier related.The failure was caused by manufacturing components out of specifications.The event involved a product problem indicating a non-conformity which is investigated and addressed through a nc and a capa.No further action is required at this time.Product surveillance will continue to monitor complaints of this type for adverse trends.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
 
Event or Problem Description
It was reported that patient presented to surgeon's office with a broken long strut.This was not a routine followup visit.Causes or contributors to strut breakage currently unknown.A wire was utilized to hold the strut in place until it can be replaced (b)(6)2018 as the surgeon is away until that date.Rep won't be able to get any additional information until on or after (b)(6)2018 when he can confer with the surgeon, but expects the device to be available for return after that date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEXAPOD STRUT, LONG HOFFMANN LRF 183MM-295MM
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key7853051
Report Number0008031020-2018-00588
Device Sequence Number11080591
Product Code KTT
UDI-Device Identifier07613327294200
UDI-Public07613327294200
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K163656
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number49350040
Device Lot NumberY14369
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2018
Initial Date Received by Manufacturer 08/09/2018
Supplement Date Received by Manufacturer11/09/2018
Initial Report FDA Received Date09/06/2018
Supplement Report FDA Received Date12/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-