| Model Number |
98100 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation conclusion: the customer's complaint was not replicated during in-house testing with retains of device lot w64237b.No issues with d-dimer recovery were observed; lot performed properly.Manufacturing batch records for the lot were reviewed, lot met release specifications.Based on the information available, there is no indication of a product deficiency and no corrective action is required.
|
| |
|
Event or Problem Description
|
|
The customer reported variability between results produced on the triage d-dimer system.Customer stated on (b)(6) 2018 a patient presented to the clinic with "pain in lower left extremity (clot)".Patient sample was tested on triage and produced d-dimer results of 202ng/ml, 163ng/ml, and 571ng/ml.The specific time of testing was not provided but customer stated testing was all done on the same day.Clinics reference range for triage d-dimer is <400ng/ml.Customer stated no other lab tests were performed and customer did not know patient diagnosis/outcome, treatment or medical history.
|
| |
|
Search Alerts/Recalls
|