| Model Number |
5703690 |
| Medical Device Problem Codes |
Break (1069); Corroded (1131)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device is pending return to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation for intubating the patient, the tunnel needle allegedly had rust on it.There was no reported patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: manufacturing records were reviewed (dhr, mrr¿s, scrap and mfg.Process changes) and there were not found evidence that the failure mode reported in this complaint is caused by the mfg.Process.Investigation summary: one hemosplit 14.5cm 19cm catheter kit was returned for evaluation.The kit was returned opened.It was noted that the tunneler possessed a reddish-brown residue on its surface.As the residue is consistent with rust, the investigation is confirmed for the reported rust.As the kit was received open, the definitive root cause could not be determined based upon available information.It is unknown whether procedural issues contributed to the reported event.Potential contributing factors include storage conditions and storage of the kit while unsealed.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiry date: 02/2019).
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation for intubating the patient, the tunnel needle allegedly had rust on it.There was no reported patient contact.Another device was used to complete the procedure.
|
| |
|