| Model Number |
S17-45 |
| Medical Device Problem Codes |
Break (1069); Material Fragmentation (1261)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Vascular Dissection (3160); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
08/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in the event will not be returned for evaluation, they remain implanted in the patient.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
During an initial procedure the marker band on the s17-45 sheath sheared off while removing the sheath.The issue was not found until after the vessel was closed at the end of the procedure.It was visualized under fluoroscopy but could not determine if it was in the vessel or in the subcutaneous layer.The physician decided to cut down in the patients right groin to explore.The marker band was found to partially be in and out of the vessel.The physician removed the marker band successfully.The patient is doing well post operatively.There has been no additional patient sequelae reported at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
A clinical evaluation of the reported event could not be completed due to a lack of receipt of relevant medical records and/or imaging; requests were made and an inadequate response was received.As such, event determination, off-label conditions, related patient harms, and patient disposition could not be independently assessed.The manufacturing lot review confirmed all devices met specifications prior to release.These types of events will be monitored and trended as part of the quality system.Patient code - remove 3165.Conclusion code - remove 11.
|
| |
|
Search Alerts/Recalls
|