| Model Number |
MMT-523NAH |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The patient¿s blood glucose level was unknown at the time of the incident.The customer also reported that the insulin was squirting during manual prime process.The device is not expected to be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received compromised force sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform basic occlusion test, occlusion test, prime test, excessive no delivery test due to compromised force sensor system alarm.However, insulin pump passed rewind test and displacement test.
|
| |
|