• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA AMISTEM-H PROXIMAL COATING STD STEM SIZE 5; CEMENTLESS HIP STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA AMISTEM-H PROXIMAL COATING STD STEM SIZE 5; CEMENTLESS HIP STEM Back to Search Results
Catalog Number 01.18.175
Health Effect - Clinical Code No Code Available (3191)
Date of Event 08/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 04 september 2018.Lot 180062: (b)(4) items manufactured and released on 08 march 2018 expiration date: 2023-04-26 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Other device involved: biolox delta ceramic ball head 12/14 ø 32 size m 0 reference 01.29.205.Lot 178996: (b)(4) items manufactured and released on 11 may 2018.Expiration date: 2023-04-29 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Visual inspection performed by r&d project manager on 29 august 2018.Femoral stem and head do not show any particular sign to report except for the extraction procedure marks and scratches.Failure cause due to leg length discrepancy.
 
Event or Problem Description
The patient came in complaining of a leg length discrepancy 15 days after primary surgery.The surgeon revised the head and stem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMISTEM-H PROXIMAL COATING STD STEM SIZE 5
Common Device Name
CEMENTLESS HIP STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
MDR Report Key7854961
Report Number3005180920-2018-00664
Device Sequence Number16761452
Product Code LZO
UDI-Device Identifier07630030804403
UDI-Public07630030804403
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K161635
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/26/2023
Device Catalogue Number01.18.175
Device Lot Number180062
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2018
Initial Report FDA Received Date09/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-