DEPUY ORTHOPAEDICS, INC. 1818910  ALTRX +4 10D 36IDX62OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 122136162
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 08/13/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the original competitor bipolar implanted 8 years ago.On (b)(6) 2018, converted to a total with pinnacle cup/liner construct.Competitor femoral implants left implanted.Patient has dislocated.Removed well fixed competitor's stem/head construct due to retroversion.Doi: (b)(6) 2018, dor: (b)(6) 2018, left hip.
 
Additional Manufacturer Narrative
Investigation summary: no device associated with this report was received for examination.Examination of the provided x-ray image confirms the reported dislocation event.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRX +4 10D 36IDX62OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key7854995
Report Number1818910-2018-68706
Device Sequence Number1986774
Product Code LPH
UDI-Device Identifier10603295016458
UDI-Public10603295016458
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K062148
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Catalogue Number122136162
Device Lot NumberH62191
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/13/2018
Supplement Date Received by Manufacturer10/04/2018
Initial Report FDA Received Date09/07/2018
Supplement Report FDA Received Date10/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-