COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 840
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during patient use the 840 ventilator generated a high alarm, the screen went blank and ventilation stopped.The patient was manually ventilated via an ambu bag and placed on an alternate ventilator.There was no patient harm or injury as a result of the reported event.The customer's respiratory therapist evaluated the ventilator.Upon turning the ventilator on, the ventilator alarmed and message showed "needs service".
 
Additional Manufacturer Narrative
Correction to the aware date on initial mdr.The date should have been 2018 (b)(4).Correction to evaluation codes method, result and conclusion.Device evaluation summary: the field service engineer (fse) evaluated the ventilator and found that the ventilator logged error codes.The fse replaced the gui (graphic user interface) central processing unit (cpu) pcb (printed circuit board) and updated the required software associated with the gui pcu pcb.The ventilator passed all testing per manufacturer specifications and was returned to the customer for clinical use.If the replaced part is returned for failure investigation, a supplement medwatch report with the findings will be submitted once the investigation is complete.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Device evaluation summary: the graphical user interface printed circuit board (xena ii) was returned for failure investigation.The part was visually inspected and functionally tested with no anomalies observed.Conclusion: no fault found.There was no malfunction or product deficiency identified.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key7855391
Report Number8020893-2018-00432
Device Sequence Number16631796
Product Code CBK
UDI-Device Identifier10884521076877
UDI-Public10884521076877
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K151252
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number840
Device Catalogue Number4-840120DICU-EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Initial Date Received by Manufacturer 08/16/2018
Supplement Date Received by Manufacturer09/10/2018
11/08/2018
Initial Report FDA Received Date09/07/2018
Supplement Report FDA Received Date09/10/2018
11/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight107
-
-