• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD VERIFY; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD VERIFY; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3531
Medical Device Problem Codes Failure to Deliver Energy (1211); Installation-Related Problem (2965)
Health Effect - Clinical Codes Micturition Urgency (1871); Therapeutic Effects, Unexpected (2099); Urinary Retention (2119)
Date of Event 08/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a consumer regarding a patient with an external neurostimulator.It was reported that the patient stopped feeling stimulation.The patient mentioned that he had changed between a couple of programs.Troubleshooting attempts were made and stimulation was on.It was noted that the patient was set to program 7 at 8.0.On (b)(6), the patient reported that the patient saw the message that the device was not set up properly for the patient's use, "contact your clinician.Turn off therapy/end session".The patient confirmed that the trial was working well and it was making him pee more with a good flow but now he was peeing less and had the urge to go again but nothing comes out.The patient was redirected to his health care professional.There were no further symptoms or complications reported or anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERIFY
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
MDR Report Key7855789
Report Number3004593495-2018-00845
Device Sequence Number11120590
Product Code EZW
UDI-Device Identifier00643169725072
UDI-Public00643169725072
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3531
Device Catalogue Number3531
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2018
Initial Report FDA Received Date09/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-