Catalog Number 71362935
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Information (3190)
Date of Event 08/13/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the drill broke at the spring section during drilling.The drill was engaged in the r3 shell screw hole when it broke, the broken piece was retrieved.
 
Additional Manufacturer Narrative
The associated complaint device was not returned.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the listed part did not reveal additional complaints for the listed batch.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Additional Manufacturer Narrative
The flexible drill was returned and evaluated.A visual inspection found that the flexible shaft has failed.The flexible shaft can fail if the device is over-torqued during operation, putting forces on the shaft that exceed the strength of the material.We recommend that all reusable instruments be routinely inspected for wear and damage so they can be replaced as necessary.No additional complaints were found during the complaint history review.Reviewing the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Based on this investigation, the need for corrective action is not indicated.No additional actions are being taken at this time; however we will continue to monitor for future complaints and investigate further as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE DRILL 35MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7855983
Report Number1020279-2018-01749
Device Sequence Number19121589
Product Code JDI
UDI-Device Identifier03596010464668
UDI-Public03596010464668
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71362935
Device Lot Number17JSC0008
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date09/07/2018
Supplement Report FDA Received Date11/01/2018
01/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-