TERUMO MEDICAL CORPORATION PINNACLE PRECISION ACCESS SYSTEM; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Codes Difficult to Flush (1251); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device was not returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.For this reason, evaluation code has been referenced in the conclusions.A review of the device history record of the product code/lot# combination was conducted with no findings.
 
Event or Problem Description
The user facility reported that while using a precision device the sheath would not flush in the patient.The sheath was removed and another hole was noticed where saline was spraying out.They used another precision and it worked fine.The patient was reported to be fine.The procedure outcome was a success.Additional information was received on 8/24/2018: the procedure was a pedal case.The saline was spraying through a sheath.Additional information was received on 8/27/2018: there was a tiny hole on the lateral portion of the sheath.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the completed investigation results.The evaluation of the actual device could not be conducted due to the device not being returned.With no device return, the exact cause of the reported event cannot be definitively determined based on the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE PRECISION ACCESS SYSTEM
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
MDR Report Key7856069
Report Number1118880-2018-00107
Device Sequence Number11379857
Product Code DYB
UDI-Device Identifier00389701010762
UDI-Public00389701010762
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K111606
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model NumberN/A
Device Catalogue Number70-4160
Device Lot NumberWD21
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/22/2018
Supplement Date Received by Manufacturer10/03/2018
Initial Report FDA Received Date09/07/2018
Supplement Report FDA Received Date10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-