| Model Number |
H74939401153500 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.The returned product consisted of a wolverine balloon catheter.A visual examination identified that the balloon of the device was tightly wrapped and had not been subjected to positive pressure.A visual and microscopic examination observed no damage to the tip or blades.All blades were present and fully bonded to the balloon surface.During the product analysis the investigator was unable to load the device on to a boston scientific 0.014" guidewire due to damage to the guidewire lumen.The guidewire used by the customer was not returned for analysis.A microscopic examination identified that the inner guidewire lumen of the shaft was stretched and bunched approximately 20mm proximal of the proximal balloon bond.A pinhole on the outer shaft was also identified located approximately 21mm proximal of the proximal balloon bond.A visual and tactile examination identified multiple kinks along the length of the hypotube of the device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on analysis completed 16 aug 2018.It was reported that a guidewire was unable to pass through the guidewire lumen.They used another of the same device to complete the procedure.It never went into the patient's body.However returned device analysis revealed there was a hole in the outer shaft of the device.
|
| |
|
Search Alerts/Recalls
|