DEPUY MITEK LLC US HEALIX BR 3.4MM; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 222343
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/08/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was received and evaluated.Visual observation confirms the anchor is broken in half.Anchor breakages are typically associated with off axis insertion, levering during insertion, hard bone quality or using incorrect instrumentation for preparing bone hole.It was reported that the surgeon did not tap and was off angle when inserting the anchor.As stated in the product ifu; ¿inserting the awl/tap or drill bit less than the specified depth, axial misalignment or levering with the anchor upon insertion may result in anchor fracture¿.A review into the depuy synthes mitek complaints system revealed no other complaints of any kind for this lot of devices that were released to distribution.This failure can be attributed to user technique; failure to follow instructions for use.No corrective action is required and no further action is warranted at this time.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.A review of the device history record indicated that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.Udi: (b)(4).This report is being filed from the (b)(4) as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported by the sales rep that during a shoulder procedure when inserting the customer's healix br 3.4mm anchor, it was observed that the device broke in half.The sales rep stated that the surgeon removed the broken anchor with no issues.The sales rep stated that there was no debris left in the patient.The sales rep reported that the surgeon completed the procedure with another like device by tapping and using the same bone hole with no patient consequences but there was a two minute delay.The sales rep stated that he believes the anchor broke due to the surgeon was off angle and did not tap when inserting the anchor.The sales rep stated that the patient's bone quality was average and good bone.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALIX BR 3.4MM
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7856617
Report Number1221934-2018-51135
Device Sequence Number13938227
Product Code MAI
UDI-Device Identifier10886705023820
UDI-Public10886705023820
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K133794
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2017
Device Catalogue Number222343
Device Lot Number3801082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2016
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/08/2015
Initial Report FDA Received Date09/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-