Device is a combination product.(b)(4).Device evaluated by mfr: the synergy ii us mr 3.50 x 12 mm stent delivery system (sds) was returned for analysis.A visual examination of the stent found no issues on the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.The tip was visually and microscopically examined and no signs of damage were noted.A visual and tactile examination found a hypotube break at 195mm distal from the distal end of the strain relief.Multiple hypotube kinks were also noted along several locations of the hypotube shaft during device examination.A review of the manufacturing process was carried out as part of the analysis.During the manufacturing process the returned device was checked for leaks at the vacuum decay test (vdt) work-step and no anomalies were noted.A visual and tactile examination of the outer and mid-shaft section found no issues with the extrusion shaft.The bi-component bond showed no signs of damage or strain.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion was unable to be determined.(b)(4).
|