| Model Number |
MSERIES BI-PHASIC |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a patient, the device inappropriately shut down.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the reported malfunction was not replicated or confirmed.The device was put through extensive testing which included functional testing without duplicating the malfunction.The device's upper housing was replaced as a precaution.It's important to mention that the batteries used during the time of the event was not returned for evaluation as part of the investigation.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|