Consumer reported complaint for low blood glucose test results.Humana representative, with customer on the line, is calling on behalf of the customer.The customer is concerned with tests results of 40 mg/dl fasting and meter to lab comparison results of 104 mg/dl non-fasting using truemetrix air meter and 285 mg/dl non-fasting from doctor's lab.The customer's expected fasting blood glucose test result range is 90 - 100 mg/dl.At the time of the call, the customer felt well and did not report any symptoms.Customer stated that in (b)(6) (customer did not remember exact date) had felt dizzy and lightheaded.Customer had performed a blood test fasting using the truemetrix air meter and obtained a result of 60 mg/dl.Customer stated she had gone to the er.Customer was diagnosed with hypoglycemia and was given orange juice.Customer did not remember her blood glucose test result when at the er.The er doctor had advised the customer to stop taking metformin and victoza.Customer was at the er for two hours before being released the same day.Customer had continued to use the truemetrix air meter.During the call on (b)(6) 2018, a back to back blood test was performed by the customer fasting and produced test results of 92 mg/dl and 97 mg/dl using truemetrix air meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 07/17/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: result 1 :40 mg/dl date: (b)(6) 2018 time:3:00 pm fasting, result 2 :104mg/dl date: (b)(6) 2018 time:7:10pm fasting, result 3 :100mg/dl date: (b)(6) 2018 time:7:15am fasting, result 4 : 90mg/dl date: (b)(6) 2018 time: 7:25am fasting, result 5 :100mg/dl date: (b)(6) 2018 time:4:50am fasting.
|