• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY; MULTIFOCAL IOLS Back to Search Results
Model Number ZXR00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/21/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zxr00 17.0 diopter intraocular lens (iol) was implanted in the patient's right eye (od) on (b)(6) 2018.It was later explanted on (b)(6) 2018 due to a negative dysphotopsia.The replacement lens was a zcb00 17.0 diopter lens.There was no incision enlargement, vitrectomy, or sutures used.Reportedly, the patient is doing good.No additional information was provided.Patient has bilateral implants.This report represents the right eye.
 
Additional Manufacturer Narrative
Device available for evaluation: yes.Returned to manufacturer on 9/10/2018.Device returned to manufacturer: yes.Device evaluation: visual inspection of the returned sample shows the lens is cut in pieces most probably to make explant possible.Considering the condition of the lens, additional analysis is not possible.Based on the visual evaluation of the returned unit it could not be determined that the cause of the issue reported is related to the manufacturing process.A product quality deficiency could not be confirmed.Manufacturing records review: during the manufacturing record review of the production order for this serial number, no product deficiency was identified.The documentation shows that the production order was manufactured according to specifications.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Labeling review: the labeling review was completed and revealed that the directions for use (dfu) provide the customer with proper usage instructions and guidelines.No labeling changes are required.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7857117
Report Number9614546-2018-00869
Device Sequence Number13735580
Product Code POE
UDI-Device Identifier05050474579033
UDI-Public(01)05050474579033(17)220816
Combination Product (Y/N)Y
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/16/2022
Device Model NumberZXR00
Device Catalogue NumberZXR00U0170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/13/2018
Supplement Date Received by Manufacturer10/19/2018
10/25/2020
Initial Report FDA Received Date09/07/2018
Supplement Report FDA Received Date11/15/2018
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Date Report Sent to FDA01/01/2005
-
-